Hydrophone Calibration with Temperature
GeoSpectrum’s high-speed hydrophone calibrator can be equipped to measure voltage sensitivity and capacitance as a function of temperature. In the standard form of the equipment, the operator places the calibration chamber within a temperature-controlled cabinet. After the temperature of the calibration chamber has reached the new temperature, the operator activates the calibration software. Optionally, the computer can control the temperature chamber and calibrate the hydrophones at temperatures specified by the operator.

Typical graphs of capacitance and voltage sensitivity as a function of temperature
Calibrations as a function of temperature can be combined with calibrations as a function of static pressure and signal amplitude. Distortion can be measured at any combination of temperature, depth, and signal amplitude.
Parameters measured
Temperature range
Accuracy
Repeatability
Cycle time for calibration
Stabilization time at each temperature
Quantity calibrated per cycle
Measurement medium
Ambient pressure
Measurement frequency
Dynamic pressure amplitude (rms)
Data recorded
Calibration chamber size (H x W x D)
Usable calibration volume |
Voltage sensitivity
Capacitance
-20 °C to +80 °C
± 0.3 dBV (NIST traceable)
± 0.1 dBV
30 seconds
4 hours minimum
10
Air
Atmospheric, standard
26 Hz
Adjustable, but typically 1 mbar
Voltage sensitivity
Capacitance
Charge sensitivity (calculated)
Temperature
Serial number
Time of calibration
14 x 15 x 18 cm, standard
1000 cc, standard
|
|