Products
Calibration Equipment
Hydrophones Sound Projectors Calibration Equipment
Hydrophone Calibration at Great Depth (up to 2700m)
This calibrator is capable of calibrating eight hydrophones at a time at depths up 2700 m. The unknown and reference hydrophones are placed in a hydraulic pressure vessel that contains electrically-shielded sound projectors in the bottom. After the vessel has been sealed and bled of air, the vessel is pressurized to the chosen depth, and the calibration software activated.
This equipment comes in two versions, both of which produce calibrations of the same accuracy. In the simpler version, the operator manually changes the pressure with a hand-operated hydraulic pump and manually activates the calibration software. In the automatic version, the operator enters a pressure-versus-time profile and the times at which calibrations are to be performed. The software controls a hydraulic pump and electric valves to produce the specified pressure profile and calibrates the hydrophones at the specified times. The calibration results are written to a file in both versions.

Top view of pressure vessel

Screen shot of the automatic pressure-control/calibration software. The blue line in the upper left graph is the user-specified pressure profile; the red line is the pressure achieved by the system. The red dots are the times at which calibrations are performed. The graph on the upper right shows a time trace of the reference and unknown hydrophone signals. The graph immediately below shows the FFT of the hydrophone signals
Calibration System
Parameters measured
Absolute accuracy of voltage sensitivity
Relative accuracy of voltage sensitivity
(one hydrophone to another)
Measurement frequency
Cycle time for calibration
Number of unknown hydrophones
Number of calibration points
Data recorded
Pressure Control System
Depth range
Measurement medium
Number of pressure steps in the pressure profile
Maximum time of the pressure profile
Maximum rate of change of pressure
Usable calibration chamber dimensions
|
Voltage sensitivity
Capacitance
± 2.0 dBV
± 0.1 dBV
15 to 200 Hz
30 seconds
8
Unlimited
Voltage sensitivity
Capacitance
Charge sensitivity (calculated)
Depth
Temperature
Time of calibration
30 to 2700 m
Hydraulic fluid
Unlimited
Unlimited
Equivalent to 10 m/s
30 cm high, 11 cm diameter
(other sizes available)
|
|